Heath-consultants Gas Delivery Unit (GDUnet) Manuel d'utilisateur Page 43

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 109
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 42
net
3-15
3.3.7 GDUnet Selection
 net must have the

 net registered on the network, in the ‘Send
To’ section, click the drop down arrow and select the required
net, as illustrated in Fig. 3-17.
Fig. 3-17 GDUnet Selection
3.3.8 Test or Calibrate Selection
‘Send To’ also contains the selection to Te st or Calibrate’
instruments as the required process.
The ‘Test’ process is selected to simply apply gas then
record instrument status.
This process does not update instrument calibration
settings in any way.
The ‘Calibrate’ process is selected to perform a full
calibration of the instrument. Gases are applied and the
instrument settings are automatically adjusted until the
instrument reading is the same as each gas applied.
This process updates stored data including new ‘cal-due’
date.
To select ‘Test’ or ‘Calibrate’ option, select the appropriate ‘radio’
button, highlighted in Fig.3-18.
Vue de la page 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 108 109

Commentaires sur ces manuels

Pas de commentaire